Quality assurance of gFOBT testing
Whilst an immunochemical test is recommended, programmes that adopt a traditional guaiac
test need to apply additional laboratory quality procedures. To minimise variability and error
associated with visual test reading, including manual results input, the following procedures
should be considered
o Use of appropriate temperature for artificial lighting and neutral-coloured walls in the
reading laboratory;
o Use of a national laboratory training programme to prosper consistency of interpretation;
o A blinded internal QC check each day for each analyst prior to commencing testing;
o Adoption of a monitoring programme to identify operator related analytical performance
(e.g. positivity variability and bias); and
o Double entry of test results
Quality assurance of iFOBT testing
Consistency in analytical performance must be assured by the adoption and application of rigorous
quality assurance procedures. Manufacturer’s Instructions for Use must be followed. Laboratories
should perform daily checks of analytical accuracy and precision across the measurement
range with particular emphasis at the selected cut-off limit. Rigorous procedures need to
be agreed and adopted on how internal quality control data is interpreted and how the laboratory
responds to unsatisfactory results. Performance data, both internal quality control and
external quality assessment data, should be shared and reviewed by a Quality Assurance team
working across the programme. Sufficient instrumentation should be available to avoid delays in
analysis due to instrument failure or maintenance procedures
External quality assessment
A European external quality assessment scheme should be developed to facilitate Europe-wide
quality assurance of occult blood testing and enhance the reproducibility of testing within and
between countries providing population screening
Outcome monitoring
All aspects of laboratory performance in respect of the screening test should be part of a rigorous
quality assurance system. Uptake, undelivered mail, time from collection to analysis, analytical
performance (internal QC and external QA), positivity rates, lost & spoilt kits and technical
failure rate, technician performance variability and bias should each be subject to rigorous
monitoring
Quality of information
The proportion of unacceptable tests received for measurement is influenced by the ease of use
of the test kit and the quality of the instructions for use. This proportion should not exceed 3%
of all kits received; less than 1% is desirable
No comments:
Post a Comment